i. Select direct low frequency readout using meter Function 13. This
is done by selecting F13 on the meters front panel and setting the
instrument to then BC.
ii. Set BANDWIDTH to 15 kHz (NB)
iii. Turn FSM RF Amplifier OFF
iv. Set AUDIO to AM (This setting does not affect measurements in
any way. It only affects the detector driving the internal speaker.)
v. If not in the dBuV read out mode, set measurement units to dBuV
using meter function F14 (switches between dBm and dBuV) and
confirm signal strength is within range (+90dBuV to +10dBuV) (-
17dBm to –97dBm) of field strength meter. If the signal to be
measured is larger than –17dBm +90dBuV the SIGNAL LEVEL
readout will display "> -17 dBm 90dBuV " with the ">" sign
blinking. A calibrated 30 dB, 50 ohm coaxial pad may be installed
at input to block converter if signal strength is too high.
vi. Set measurement units to dBuV/m using meter function F15 (F15
activates inclusion of Antenna Factors from the resident table
within the meter. This resident table is unique to your calibrated
antenna and may be programmed into your meter using the
supplied Load_Tab.exe application.)
vii. Read field strength directly in dBuV/m on Field Strength Meter
front panel.
b. If making measurement using the PC application:
i. Set BANDWIDTH to 15 kHz (NB)
ii. Turn FSM RF Amplifier OFF
iii. Set AUDIO to AM (This setting does not affect measurements in
any way. It only affects the detector driving the internal speaker.)
iv. Using the FSM front panel, confirm that the meter reads in the
dBuV. If not in the dBuV read out mode, set measurement units to
dBuV using meter function F14 (switches between dBm and
dBuV) and confirm signal strength is within range (+90dBuV to
+10dBuV) of field strength meter. If the signal to be measured is
larger than +90dBuV the SIGNAL LEVEL readout will display ">
90dBuV " with the ">" sign blinking. A calibrated 30 dB, 50 ohm
coaxial pad may be installed at input to block converter if signal
strength is too high.
v. Place the FSM in Remote mode by selecting function 93 (F93) on
the meter front panel.
vi. Connect the meter to the PC serial port
vii. Launch the Application by double clicking on the desktop icon.
viii. Enter antenna factor for sum of antenna and cable and pad (if used)
ix. Read field strength directly in dBuV/m on the PC display.
4. For mobile measurements using Z Technology GPS aware software:
a. As described in an earlier section, set frequency, bandwidth, block
converter offset, antenna factor, and readout mode on the FSM front panel.